SpecificationsTeledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
29
2.2.
EPA EQUIVALENCY DESIGNATION
Note
GFC7001TM: EPA equivalency does not apply to this model.
TAI’s Model GFC7001T, Carbon Monoxide Analyzer, is designated as Reference
Method Number RFCA-1093-093 as defined in 40 CFR Part 53, when operated under
the following conditions:
•
Range: Any range from 10 ppm to 50 ppm
•
Ambient temperature range of 10 to 40
°
C
•
Sample filter: Equipped with 5-micron PTFE filter element in the internal
filter assembly
•
Software settings:
Dilution factor
1.0
AutoCal
ON or OFF
Dynamic Zero
ON or OFF
Dynamic Span
OFF
Dual Range
ON or OFF
Auto Range
ON or OFF
Temp/Pres Compensation
ON
Under the designation, the analyzer may be operated with or without the following
options:
•
Rack mount with slides
•
Rack mount without slides, ears only
•
Zero/span valve options
•
Option 50A – Sample/Cal valves, or
•
Option 50B – Sample/Cal valves with span shutoff & flow control
•
Internal zero/span (IZS) option with either:
•
Option 51A – Sample/Cal valves, or
•
Option 51C – Sample/Cal valves with span shutoff & flow control
•
4-20mA, isolated output
2.3.
APPROVALS AND CERTIFICATIONS
The TAI Model GFC7001T/GFC7001TM analyzer was tested and certified for Safety
and Electromagnetic Compatibility (EMC). This section presents the compliance
statements for those requirements and directives.
Accuracy
<± (0.02% CO
2
+ 2% of reading)
Linearity
<± 0.1% CO
2
Temperature Coefficient
<± 0.01% CO
2
/°C
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
Содержание GFC7001T
Страница 100: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 100 ...
Страница 107: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 107 ...
Страница 118: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 118 ...
Страница 136: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 136 This page intentionally left blank ...
Страница 195: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 195 ...
Страница 207: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 207 ...
Страница 211: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 211 ...