Calibration Procedures Teledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
208
9.7.1.3.
ACTIVATE O2 SENSOR STABILITY FUNCTION
To change the stability test function from CO concentration to the O
2
sensor output,
press:
Note
Use the same procedure to reset the STB test function to CO when the O2
calibration procedure is complete.