EPA Calibration Protocol Teledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
223
SPAN CAL M
A1:CONC1=50 PPM CO = XXXX
< TST TST >
ZERO
SPAN
CONC
EXIT
Wait until
STABIL
falls
below
0.2 PPM
(for M300E).
This may take
several minutes.
SAMPLE
COSTB=XXXX PPB
CO=XXXX
< TST TST > CAL
CALZ
CALS
SETUP
Record the CO
reading as
displayed on the
instrument’s front
panel.
Press
EXIT
to
Return to the
Main
SAMPLE
Display.
ACTION:
Allow Calibration Gas diluted to
proper concentration for
Midpoint N+1 to enter the sample
port.
ACTION:
Allow calibration gas diluted to proper concentration for
Midpoint N to enter the sample port
SAMPLE A1:CONC1=50 PPM CO = XXXX
< TST
TST
>
CAL
SETUP
SAMPLE
CO STB=XXXX PPB
CO=XXXX
< TST TST >
CAL
SETUP
Set the Display to show the
COSTB
test function.
This function calculates the
stability of the CO
measurement.
Plot the analyzer responses versus the corresponding calculated concentrations to obtain
a calibration relationship. Determine the best-fit straight line (y = mx + b) determined
by the method of least squares.
After the best-fit line has been drawn, determine whether the analyzer response is linear.
To be considered linear, no calibration point should differ from the best-fit line by more
than
±
2% of full scale.
If carried out carefully, the checks described in this section will provide reasonable
confidence that the GFC7001T is operating properly. Checks should be carried out at
least every 3 months as the possibility of malfunction is always present.
If the linearity error is excessive and cannot be attributed to outside causes, check the
GFC7001T system for:
•
Sample pressure higher than ambient – pressurized sample gas
•
Leaks
•
Correct flow
•
Miscalibrated span gas tanks or bad zero gas
•
Miscalibrated sample pressure transducer
Содержание GFC7001T
Страница 100: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 100 ...
Страница 107: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 107 ...
Страница 118: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 118 ...
Страница 136: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 136 This page intentionally left blank ...
Страница 195: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 195 ...
Страница 207: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 207 ...
Страница 211: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 211 ...