Communications Setup and OperationTeledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
141
Figure 6-3:
COMM – COM1 Test Port
6.3.
RS-232
The RS232 and COM2 communications (COMM) ports operate on the RS-232 protocol
(default configuration). Possible configurations for these two COMM ports are
summarized as follows:
•
RS232 port can also be configured to operate in single or RS-232 Multidrop mode
(Option 62); refer to Sections 3.3.1.9 and 5.7.1.
•
COM2 port can be left in its default configuration for standard RS-232 operation
including multidrop, or it can be reconfigured for half-duplex RS-485 operation
(please contact the factory for this configuration).
Note that when the rear panel
COM2
port is in use, except for multidrop
communication, the rear panel USB port cannot be used. (Alternatively, when the USB
port is enabled,
COM2
port cannot be used except for multidrop).
A Code-Activated Switch (CAS), can also be used on either port to connect typically
between 2 and 16 send/receive instruments (host computer(s) printers, data loggers,
analyzers, monitors, calibrators, etc.) into one communications hub. Contact TAI Sales
for more information on CAS systems.
To assist in properly connecting the serial ports to either a computer or a modem, there
are activity indicators just above the RS-232 port. Once a cable is connected between
the analyzer and a computer or modem, both the red and green LEDs should be on.
•
If the lights are not lit, use small switch on the rear panel to switch it between DTE
and DCE modes.
Select which
COMM port to
test.
SETUP X.X
COMMUNICATIONS MENU
ID INET
COM1
COM2
EXIT
SETUP X.X
COM1 : TEST PORT
<SET
TEST
EXIT
SETUP X.X
TRANSMITTING TO COM1
<SET
TEST
EXIT
SETUP X.X
COM1 MODE:0
SET>
EDIT
EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET
SET>
EDIT
EXIT
EXIT returns to
COMM menu
Test runs
automatically
Содержание GFC7001T
Страница 100: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 100 ...
Страница 107: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 107 ...
Страница 118: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 118 ...
Страница 136: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 136 This page intentionally left blank ...
Страница 195: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 195 ...
Страница 207: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 207 ...
Страница 211: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 211 ...