SetupTeledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
110
To access and navigate the
VARS
menu, use the following button sequence.
SETUP X.X 4) CLOCK_ADJUST=0 Sec/Day
PREV
NEXT
JUMP
EDIT ENTR EXIT
SETUP X.X CLOCK_ADJUST=0 Sec/Day
+
0
0
ENTR EXIT
Enter sign and number of
seconds per day the clock
gains (-) or loses(+).
SETUP X.X
DYN_SPAN=OFF
OFF
ENTR EXIT
SETUP X.X
DYN_ZERO=OFF
OFF
ENTR EXIT
Toggle this button to turn
the Dynamic Zero
calibration feature
ON/
OFF.
Concentration display
continuously cycles
through all gasses.
SAMPLE
RANGE=500.0 PPM CO= XXXX
<TST
TST> CAL
SETUP
SETUP X.X
0) DAS_HOLD_OFF=15.0 Minutes
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
Toggle these
buttons to enter
the correct
PASSWORD.
In all cases:
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
SETUP X.X
2) DYN_ZERO=OFF
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
3) DYN_SPAN=OFF
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
DAS_HOLD_OFF=15.0 Minutes
1
5
.0
ENTR EXIT
Toggle these buttons to set
the iDAS HOLDOFF time
period in minutes
(MAX = 20 minutes).
Toggle this button to turn
the Dynamic Span
calibration feature
ON/
OFF.
SETUP X.X
1) CONC_PRECISION=AUTO
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
CONC_PRECISION=AUTO
AUTO
1
2
3
4
ENTR EXIT
SETUP X.X
5) STABIL_GAS=CO
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
STABIL_GAS=O2
CO
CO2
O2
ENTR EXIT
Use these buttons to select
which gas will be reported
by the sTABIL test
function.
(O
2
is only available if the
optional O
2
sensor is
installed)
Use these buttons to select
the precision of the o33
concentration display.
Press
NEXT
for additional
VARS; press
NEXT
or
PREV
to move back and
forth throughout the list of
VARS.
5.9.
SETUP
Æ
DIAG: DIAGNOSTICS FUNCTIONS
A series of diagnostic tools is grouped together under the
SETUP
Æ
MORE
Æ
DIAG
menu, as these parameters are dependent on firmware revision (see Appendix A). These
tools can be used in a variety of troubleshooting and diagnostic procedures and are
Содержание GFC7001T
Страница 100: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 100 ...
Страница 107: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 107 ...
Страница 118: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 118 ...
Страница 136: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 136 This page intentionally left blank ...
Страница 195: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 195 ...
Страница 207: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 207 ...
Страница 211: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 211 ...