Calibration Procedures Teledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
194
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
For this Initial Calibration it is important to independently verify the
PRECISE CO Concentration Value of the SPAN gas.
If the source of the Span Gas is from a Calibrated Bottle, use the exact
concentration value printed on the bottle.
9.3.3.2.
ZERO/SPAN POINT CALIBRATION PROCEDURE
The zero and cal operations are initiated directly and independently with dedicated
buttons (
CALZ
&
CALS
).