Troubleshooting and ServiceTeledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
269
7.
Carefully remove the opto-pickup printed circuit assembly.
Opto-Pickup
Figure 12-16: Removing the Opto-Pickup Assembly
8.
Remove the three (3) screws holding the GFC Wheel motor/heat sink assembly to
the GFC Wheel housing.
9.
Carefully remove the GFC Wheel motor/heat sink assembly from the GFC Wheel
housing.
GFC WHEEL HOUSING
Figure 12-17: Removing the GFC Wheel Housing
Содержание GFC7001T
Страница 100: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 100 ...
Страница 107: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 107 ...
Страница 118: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 118 ...
Страница 136: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 136 This page intentionally left blank ...
Страница 195: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 195 ...
Страница 207: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 207 ...
Страница 211: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 211 ...