14.4.2.5.
PACKING COMPONENTS FOR RETURN TO TELEDYNE CUSTOMER
SERVICE
Always pack electronic components and assemblies to be sent to Teledyne Customer
Service in anti-ESD bins, tubes or bags.
CAUTION
ESD Hazard
•
DO NOT
use pink-poly bags.
•
NEVER
allow any standard plastic packaging materials to touch the
electronic component/assembly directly.
•
This includes
,
but is not limited to, plastic bubble-pack, Styrofoam
peanuts, open cell foam, closed cell foam, and adhesive tape.
•
DO NOT use standard adhesive tape as a sealer. Use ONLY anti-ESD tape.
Never carry the component or assembly without placing it in an anti-ESD bag or bin.
1.
Before using the bag or container allow any surface charges on it to dissipate:
•
If you are at the instrument rack, hold the bag in one hand while your wrist
strap is connected to a ground point.
•
If you are at an anti-ESD workbench, lay the container down on the
conductive work surface.
•
In either case wait several seconds.
2.
Place the item in the container.
3.
Seal the container. If using a bag, fold the end over and fastening it with anti-ESD
tape.
•
Folding the open end over isolates the component(s) inside from the effects
of static fields.
•
Leaving the bag open or simply stapling it shut without folding it closed
prevents the bag from forming a complete protective envelope around the
device.
Note
If you do not already have an adequate supply of anti-ESD bags or
containers available, Teledyne Customer Service department will supply
them (see Section 11.8 for contact information).
Follow the instructions listed above for working at the instrument rack
and workstation.
Содержание GFC7001T
Страница 100: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 100 ...
Страница 107: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 107 ...
Страница 118: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 118 ...
Страница 136: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 136 This page intentionally left blank ...
Страница 195: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 195 ...
Страница 207: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 207 ...
Страница 211: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 211 ...