Falcon GC
Maintenance & Troubleshooting
Teledyne Analytical Instruments
49
2.
With glow plug wire removed, the glow plug can be removed
from the FID body. Using a 5/16” wrench loosen the glow plug
until it comes free of the FID body. Do NOT try to pull the glow
plug from the FID body. When it is fully unscrewed the glow
plug should come away freely.
3.
Place the copper washer and glow plug shim onto the new glow
plug. The copper washer should be placed on the glow plug first,
followed by the metal shim.
4.
Insert the new glow plug into the FID body and tighten ¼ turn
past snug.
5.
Re-attach glow plug wire.
Содержание FALCON GC
Страница 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Страница 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Страница 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Страница 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Страница 58: ...Maintenance Troubleshooting Falcon GC 48 Teledyne Analytical Instruments Figure 5 1 Glow Plug Wire ...
Страница 60: ...Maintenance Troubleshooting Falcon GC 50 Teledyne Analytical Instruments Figure 5 2 FID Diagram ...
Страница 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Страница 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Страница 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Страница 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Страница 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Страница 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...