Maintenance & Troubleshooting
Falcon GC
36
Teledyne Analytical Instruments
3.
Disconnect Power and Communication Cables.
i.
Grasp the colored connector and pull it to the right until it
is free from the pins.
ii.
Do NOT pull on the wires to free the connectors.
4.
Remove Detector Screws.
i.
Gently set the GC on its front so that you will have access
to screws on the back of the chassis.
ii.
Using a Phillips-Head screwdriver remove the 5 detector
mounting screws from the chassis of the GC.