Falcon GC
Maintenance & Troubleshooting
Teledyne Analytical Instruments
39
3.
Remove the SPU oven lid by unscrewing the 4 hex stand-offs
circled below.
4.
Remove the foam oven plugs by gently pulling them straight up
and out of their slots.
5.
Disconnect the column transfer tubes from the column module
using a 1/4” open ended wrench on the hex nut and a 3/16” open
ended wrench on the column end hex union. Be sure to hold the
column end hex union firmly so it doesn’t twist the column end
while loosening the nut.
Note:
If your transfer tubes are fused silica and you’re using
graphitized vespel ferrules, the ferrules can be re-used if
not badly deformed.
Содержание FALCON GC
Страница 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Страница 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Страница 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Страница 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Страница 58: ...Maintenance Troubleshooting Falcon GC 48 Teledyne Analytical Instruments Figure 5 1 Glow Plug Wire ...
Страница 60: ...Maintenance Troubleshooting Falcon GC 50 Teledyne Analytical Instruments Figure 5 2 FID Diagram ...
Страница 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Страница 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Страница 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Страница 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Страница 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Страница 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...