Maintenance & Troubleshooting
Falcon GC
26
Teledyne Analytical Instruments
4.
Select “Det1 H2 Pressure” from the dropdown menu. Then click
“Activate” to turn on the H
2
flow.
5.
Adjust flow to 15mL/min with carrier gas flow included.
i.
Connect a flow meter to the FID exhaust port. Click the
Activate button to turn on the FID H
2
flow. If the total
flow is below 15mL/min, enter a new value in the text
box and click Set New Pressure. Continue this process
until a flow of 15mL/min is found*.
ii.
Record this value.
* If the unit is hot the flow of the carrier gas must be subtracted
from the total. For example if the air and H
2
on the FID are
turned off and the flow through the flow meter is 2mL/min,
an H
2
flow of 15mL/min will give a total flow of 17mL/min.
Содержание FALCON GC
Страница 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Страница 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Страница 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Страница 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Страница 58: ...Maintenance Troubleshooting Falcon GC 48 Teledyne Analytical Instruments Figure 5 1 Glow Plug Wire ...
Страница 60: ...Maintenance Troubleshooting Falcon GC 50 Teledyne Analytical Instruments Figure 5 2 FID Diagram ...
Страница 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Страница 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Страница 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Страница 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Страница 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Страница 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...