Maintenance & Troubleshooting
Teledyne Analytical Instruments
37
5.
Separating the FID
i.
After everything has been disconnected, grasp the FID
oven body and gently slide it away from the oven. There
may be an FID installed on the other side of the GC as
well.
ii.
The FID should slide free easily. Do NOT pull on the FID
if there is resistance. Check for any connections still
attached to the FID.
iii.
Take care to preserve the insulating foam ring around the
body of the FID against the injector oven. This will need
to be put back in place upon new FID installation.
Note: When re-installing the FID, be sure to cap or plug the
sample inlet fitting so that foam debris does not enter the
end fitting.
Note: The above picture shows the FID installed on Channel 2.
Typically all IGDL units have the FID installed on channel 1
(to the left of the oven) thus the FID would be pulled the
opposite way.
Содержание FALCON GC
Страница 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Страница 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Страница 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Страница 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Страница 58: ...Maintenance Troubleshooting Falcon GC 48 Teledyne Analytical Instruments Figure 5 1 Glow Plug Wire ...
Страница 60: ...Maintenance Troubleshooting Falcon GC 50 Teledyne Analytical Instruments Figure 5 2 FID Diagram ...
Страница 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Страница 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Страница 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Страница 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Страница 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Страница 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...