Maintenance & Troubleshooting
Falcon GC
42
Teledyne Analytical Instruments
9.
With the new column module in hand, re-connect the module
power and communication cables.
10.
Insert the column module between its mounting brackets and
press it about half way down into position.
Note: Inserting it half-way rather than all of the way allows for
easier re-connection of the transfer tubes.
11.
Re-connect the transfer tubes to the column end fittings. If using
fused silica tubes, be sure to insert the fused silica into the fitting
until you “feel” it stop against the bottom of the fitting. Tighten
the hex nut with the 1/4” wrench while holding the column hex
union with a 3/16” wrench until snug. Do not over tighten.
Содержание FALCON GC
Страница 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Страница 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Страница 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Страница 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Страница 58: ...Maintenance Troubleshooting Falcon GC 48 Teledyne Analytical Instruments Figure 5 1 Glow Plug Wire ...
Страница 60: ...Maintenance Troubleshooting Falcon GC 50 Teledyne Analytical Instruments Figure 5 2 FID Diagram ...
Страница 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Страница 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Страница 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Страница 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Страница 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Страница 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...