Select “Det1 Ign Air Pressure” from the dropdown menu.
6.
Adjust the flow until it is at 65mL/min.
i.
With the FID H
2
flow off adjust the IGN Air pressure until a
flow of 65mL/min is obtained.
ii.
Record this value
7.
Select “Det1 Run Air Pressure” from the dropdown menu.
8.
Adjust the flow until it is at 150mL/min.
i.
With the FID H2 flow off adjust the pressure until a flow
of 150mL/min is reached.
ii.
Record this value.