Falcon GC
Maintenance & Troubleshooting
Teledyne Analytical Instruments
43
12.
Press the column module down into its fully seated position on
the mounting brackets. Attach the column module to the
brackets with the screws that were removed in Step 6 using a
3/32” hex driver.
13.
Fold a thin, flexible plastic strip into a “U” shape and insert it
sideways above the column end fitting inside the slot as shown.
Insert a foam plug in between the plastic ends and slide the foam
plug down snug with the column end connection.
Note: The plastic strip
acts a “shoehorn” to enable the foam plug
to slide easily and tightly against the column end
connection thereby minimizing air leakages. It is used to
protect the edges of the column plugs.
Содержание FALCON GC
Страница 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Страница 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Страница 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Страница 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Страница 58: ...Maintenance Troubleshooting Falcon GC 48 Teledyne Analytical Instruments Figure 5 1 Glow Plug Wire ...
Страница 60: ...Maintenance Troubleshooting Falcon GC 50 Teledyne Analytical Instruments Figure 5 2 FID Diagram ...
Страница 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Страница 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Страница 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Страница 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Страница 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Страница 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...