Falcon GC
Maintenance & Troubleshooting
Teledyne Analytical Instruments
47
20.
Replace the front on the unit using the 4 screws removed in Step
1.
Installation is complete. After powering up the instrument, all of the
column's information can be viewed either from the touchscreen display
in the “Column Module” screen or in under the “GC Info” tab of the
“Special Features” box in Chromperfect.
5.6 Replacing the Glow Plug
Caution:
If the unit is on and at operating temperature the
glow plug will be very hot. It should NOT be
touched without proper precautions. If the unit is
off and near room temperature the glow plug will
be at a safe temperature and can be touched
without tools or gloves.
1.
Remove the red wire that connects the glow plug to the FID
board near the base of the FID. This can be done by lightly
pulling on the rubber sleeve covering the metal end of the glow
plug wire.
Содержание FALCON GC
Страница 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Страница 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Страница 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Страница 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Страница 58: ...Maintenance Troubleshooting Falcon GC 48 Teledyne Analytical Instruments Figure 5 1 Glow Plug Wire ...
Страница 60: ...Maintenance Troubleshooting Falcon GC 50 Teledyne Analytical Instruments Figure 5 2 FID Diagram ...
Страница 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Страница 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Страница 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Страница 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Страница 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Страница 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...