![Tektronix TLA 700 Series Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/tektronix/tla-700-series/tla-700-series_installation-manual_1078203071.webp)
Restoring and Reinstalling Software
54
TLA700 Series Logic Analyzer Installation Manual
13.
Select Load Firmware from the Execute menu.
14.
Select the tla7pg2.lod file for the selected module; if necessary, navigate to
the C:\Program Files\Tektronix Pattern Generator\Firmware folder.
15.
Click OK. You will be prompted to confirm your action; click Yes.
The program will load the firmware for each module one at a time. The
process may take several minutes per module.
16.
When the upgrade is complete, exit the program.
17.
Power down the mainframe and disconnect the power cord.
18.
Remove the module from the instrument and remove the jumper from the
Flash programming pins on the rear of the module. Keep the jumper for
future updates.
19.
Locate the big label on the side of the module.
20.
Record the firmware version that is printed on the label. You will need this
information to see that the firmware version matches the label.
21.
Reinstall the module in the mainframe and power on the mainframe.
22.
After the instrument completes the power-on diagnostics, select System
Properties from the System menu.
23.
Click the module tab (for example, PG1).
24.
Verify that the firmware version for the selected module matches the version
on the label that you recorded in step 20.
25.
If the firmware versions do not match, power down the mainframe, remove
the module from the mainframe and update the label.