Restoring and Reinstalling Software
48
TLA700 Series Logic Analyzer Installation Manual
Upgrading or Restoring Firmware
You may have to upgrade the firmware on the logic analyzer or on individual
modules if the firmware version is not compatible with the TLA application
software version. This will be indicated by one or more messages (one per
module) when you start the logic analyzer. The most current firmware for the
TLA application software resides in a file on the hard disk of the logic analyzer.
Although the firmware update procedures are similar, there are some key
differences between the Tektronix Logic Analyzer products and are therefore
listed as separate procedures. If you have mixed modules within your TLA700
mainframe, review the procedures for each of the modules before beginning the
upgrade.
Use the following procedure to upgrade the firmware for the TLA7Axx/
TLA7NAx Series Logic Analyzer Modules.
1.
Note which modules require new firmware.
If any modules are missing from the System window, you can verify the
status of the modules by doing the following steps:
a.
Select System Properties from the System menu.
b.
Click the Modules tab.
c.
Note the information under the Messages column for any installed
modules. If any modules indicate invalid or similar messages, you must
update the firmware for those modules.
2.
Exit the logic analyzer application.
3.
Click Start
→
Programs
→
Tektronix Logic Analyzer
→
TLA Firmware
Loader.
4.
You may be prompted to cycle the power on the mainframe after completing
the upgrade operation. Click Yes to continue.
The logic analyzer will scan the mainframe to determine which modules are
installed and which modules can have their firmware upgraded. If you have
any older modules, you may see them listed in the bottom half of the
window. These modules cannot be updated until you install the flash jumper
on the rear of the modules.
5.
Select the modules that you want to update from the list of modules
displayed in the Supported list box near the top of the window. If you are
updating more than one module at a time, note the slot location of the
module in the selection box.
6.
Select Load Firmware from the Execute menu.
Upgrading Firmware on
TLA7Axx/TLA7NAx Logic
Analyzer Modules