![Tektronix TLA 700 Series Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/tektronix/tla-700-series/tla-700-series_installation-manual_1078203039.webp)
Basic Installation
22
TLA700 Series Logic Analyzer Installation Manual
Connect the oscilloscope probes to the Probe Compensation connector on the
front panel of the oscilloscope modules. You can then run the Calibrate Probe
function in each vertical setup page for the module.
Set up a simple pattern generator program to output a signal to the probes.
Connect an oscilloscope probe to the probe outputs and verify that the signal is
present at the probe outputs.
To check the mainframe diagnostics not covered by the TLA Application
software, run the TLA700 Mainframe Diagnostics located under the Windows
Start menu (Start/Programs/Tektronix Logic Analyzer/TLA Mainframe
Diagnostics). Exit the TLA Application before running the external diagnostics.
Backing Up User Files
Back up your user files on a regular basis. Use the Windows backup tools or
copy the files to another media. Always keep a backup copy of files that you
access on a regular basis.
Removing the Replaceable Hard Disk Drive
CAUTION.
To avoid permanently damaging the replaceable hard disk drive, do
not remove it while the instrument is powered on. Always power down the
instrument before removing the hard disk drive.
1.
Verify that the instrument is turned off.
2.
Remove the replaceable hard disk drive cartridge by depressing it to release
the latch; see Figure 16 on page 23.
3.
Pull on the replaceable hard disk drive cartridge to remove it from the
chassis.
Checking the P624x
Oscilloscope Probes
(Optional)
Checking the P647x
Pattern Generator Probes
(Optional)
Checking the TLA700
Mainframe (Optional)