Restoring and Reinstalling Software
TLA700 Series Logic Analyzer Installation Manual
49
7.
Click the proper .lod for the selected module; refer to Table 5 on page 52 if
necessary.
NOTE
. Be sure to select the proper .lod file for each module. Note the slot
number in the title bar so that you select the correct module for the .lod file.
8.
Click OK. You will be prompted to confirm your action; click Yes.
NOTE
. The program will not allow you to load firmware to an incompatible
module. For example, the program will not load oscilloscope firmware to a logic
analyzer module.
The program will load the firmware for each module one at a time. The
process may take several minutes per module.
9.
When the firmware has been loaded for each module, exit the firmware
loader program and power down the logic analyzer. You must power down
the logic analyzer to allow the TLA application to start up properly.
10.
Remove the logic analyzer module from the mainframe.
11.
Locate the big label on the side of the module.
12.
Record the firmware version that is printed on the label. You will need this
information to see that the firmware version matches the label.
13.
Reinstall the module in the mainframe and power on the mainframe.
14.
After the logic analyzer completes the power-on diagnostics, select System
Properties from the System menu.
15.
Click the module tab (for example, LA1).
16.
Verify that the firmware version for the selected module matches the version
on the label that you recorded in step 12.
17.
If the firmware versions do not match, power down the mainframe, remove
the module from the mainframe and update the label.