Maintenance
ADA400A Differential Preamplifier
55
2.
Remove the nut on the shaft of the potentiometer and replace the potentiom-
eter with the new part. Replace and tighten the nut and washer on the shaft.
3.
Resolder the wires as Figure 20 illustrates.
Brown
Red
Orange
Orange
Brown
Red
311--2533--00
311--2158--02
(Fine)
(Coarse)
Figure 20: Potentiometer Connections
After performing the procedure to access the internal components of the
ADA400A Differential Preamplifier, replace the cable assembly as follows:
1.
Remove the remaining two screws that hold the circuit board to the bottom
housing.
NOTE
. To prevent contaminating sensitive areas of the circuit board, handle the
circuit board by the edges only.
2.
Unsolder the four colored wires from the top of the circuit board and the
three ground prongs from the bottom of the circuit board.
3.
Note the color-code information on the circuit board. Replace the cable
assembly and solder the new wires.
4.
Place the circuit board and cable strain relief in the bottom housing and
reinstall hardware.
Replacing the Cable
Assembly