Performance Verification
36
ADA400A Differential Preamplifier
Table 14: Equipment Settings for X100, 100 Hz CMRR Checks
ADA400A
Setting
Upper Bandwidth
3 kHz
Oscilloscope
Setting
Vertical gain
2 mV/div
Time base
5 ms/div
Sine Wave Generator
Setting
Output
100 Hz
18.
Set ‘--’ input to
GND
.
19.
Do steps 5 to 10.
20.
Set oscilloscope to 10
V/div. Measure trace is
<
3.2 divisions wide (32
V,
90 dB).
21.
Modify the equipment settings according to Table 15.
Table 15: Equipment Settings for X10, 100 kHz CMRR Checks
ADA400A
Setting
Gain
10
Upper Bandwidth
1 MHz
Oscilloscope
Setting
Vertical gain
20 mV/div
Time base
5
s/div
Sine Wave Generator
Setting
Output
100 kHz
22.
Set ‘--’ input to
GND
.
23.
Do steps 5 to 10 but with only one 10X attenuator.
24.
Set oscilloscope to 1 mV/div. Measure trace is
<
1 division wide (1 mV,
60 dB).
25.
Modify the equipment settings according to Table 16.