Performance Verification
32
ADA400A Differential Preamplifier
Table 10: Equipment Settings for Gain Checks (Cont.)
ADA400A Control
Setting
Offset
Off
Oscilloscope Control
Setting
Channel Selection
Ch1
Vertical input impedance
1
M
Ω
Vertical gain
10 mV/div
Time base
200
s/div
Vertical Coupling
DC
Vertical Position
0 V (mid-scale)
Trigger
Edge
Trigger position
50%
Trigger source
Ch4
Acquisition mode
Hi Res
Auto Measurement
Amplitude
Bandwidth
Full
Calibration Generator Control
Setting
Output
50 mV, 1 kHz
2.
Connect a dual input connector to inputs of ADA400A.
3.
Connect a coaxial cable from calibration generator output to dual input
connector.
4.
Adjust vertical position knob on the oscilloscope to center square wave on
screen.
5.
Check DC gain accuracy is
±
2% (49 to 51 mV). (Use amplitude measure-
ment function.)
6.
Set ADA400A gain setting to 10.
7.
Set calibration generator to 0.5 V amplitude.
8.
Adjust vertical position knob on oscilloscope to center square wave on
screen.
9.
Check DC gain accuracy is
±
2% (490 mV to 510 mV).
10.
Set ADA400A gain setting to 1.
11.
Set calibration generator to 5 V amplitude.
Содержание TEKFAX ADA400A
Страница 4: ......
Страница 8: ...Table of Contents iv ADA400A Differential Preamplifier...
Страница 18: ...Getting Started 6 ADA400A Differential Preamplifier...
Страница 50: ...Performance Verification 38 ADA400A Differential Preamplifier...
Страница 52: ......
Страница 62: ...Adjustment Procedures 50 ADA400A Differential Preamplifier...
Страница 76: ...Replaceable Parts 64 ADA400A Differential Preamplifier...