Adjustment Procedures
ADA400A Differential Preamplifier
45
4.
Adjust vertical position knob on the oscilloscope to center square wave on
screen.
5.
Adjust RV62 for 5.0 divisions on screen (50 mV).
6.
Set ADA400A gain setting to 10.
7.
Set calibration generator to 0.5 V amplitude.
8.
Adjust the vertical position knob on the oscilloscope to center square wave
on screen.
9.
Adjust RV106 to 5.0 divisions.
Equipment
Required
One digitizing oscilloscope (item 1)
1.
Connect the ADA400A to Channel 1 and make the equipment settings
according to Table 19.
Table 19: Equipment Settings for Offset Adjustments
ADA400A
Setting
+ Input
Gnd
-- Input
Gnd
Gain
10
Upper Bandwidth
100 Hz
Offset
Off (remains off through out the test)
Oscilloscope Control
Setting
Channel Selection
Ch1
Vertical input impedance
1
M
Ω
Vertical gain
500
V/div
Time base
5 ms/div
Vertical Coupling
DC
Vertical Position
0 V (mid-scale)
Trigger mode
Auto
Acquisition mode
Hi Res
Bandwidth
Full
2.
Check to make sure oscilloscope is at position 0 and vertical offset 0.
Zero Offset Adjustments