Performance Verification
34
ADA400A Differential Preamplifier
6.
Adjust the Coarse knob to bring the oscilloscope trace back to the center of
the screen.
7.
Set the vertical gain on the oscilloscope to 100 mV/div.
8.
Slowly adjust the Coarse knob to bring the oscilloscope trace to the bottom
of the screen.
9.
Adjust the Fine knob fully CW.
10.
Verify that the oscilloscope trace raises at least 4 divisions on the display.
Equipment
Required
One digitizing oscilloscope (item 1)
One sine wave generator (item 2)
One dual input coupler (item 6)
One 50
Ω
terminator (item 7)
Two 10X attenuators (item 8)
Two coaxial cables (item 9)
1.
Connect the ADA400A to Ch1 of the oscilloscope and make the equipment
settings according to Table 12.
Table 12: Equipment Settings for X100, 100 kHz CMRR Checks
ADA400A
Setting
+ Input
DC
-- Input
Gnd
Gain
100
Upper Bandwidth
FULL (1 MHz)
Offset
Off
Oscilloscope
Setting
Vertical gain
2 mV/div
Time base
5
s/div
Acquisition mode
Average
Number of waveforms to average
20
Trigger source
Ch4
Sine Wave Generator
Setting
Output
100 kHz, 1 Vp-p into 50
Ω
CMRR
Содержание TEKFAX ADA400A
Страница 4: ......
Страница 8: ...Table of Contents iv ADA400A Differential Preamplifier...
Страница 18: ...Getting Started 6 ADA400A Differential Preamplifier...
Страница 50: ...Performance Verification 38 ADA400A Differential Preamplifier...
Страница 52: ......
Страница 62: ...Adjustment Procedures 50 ADA400A Differential Preamplifier...
Страница 76: ...Replaceable Parts 64 ADA400A Differential Preamplifier...