Performance Verification
ADA400A Differential Preamplifier
37
Table 16: Equipment Settings for X10, 10 kHz CMRR Checks
ADA400A
Setting
Upper Bandwidth
100 kHz
Oscilloscope
Setting
Vertical gain
20 mV/div
Time base
50
s/div
Sine Wave Generator
Setting
Output
10 kHz
26.
Set ‘--’ input to
GND
.
27.
Do steps 5 to 10 but with only one 10X attenuator.
28.
Set oscilloscope to 100
V/div. Measure trace is
<
0.56 division wide
(56
V, 85 dB).
29.
Modify the equipment settings according to Table 17.
Table 17: Equipment Settings for X10, 100 Hz CMRR Checks
ADA400A
Setting
Upper Bandwidth
3 kHz
Oscilloscope
Setting
Vertical gain
20 mV/div
Time base
5 ms/div
Sine Wave Generator
Setting
Output
100 Hz
30.
Set ‘--’ input to
GND
.
31.
Do steps 5 to 10 but with only one 10X attenuator.
32.
Set oscilloscope to 100
V/div. Measure trace is
<
0.56 division wide
(56
V, 85 dB).
33.
Disconnect setup.
This completes the performance verification procedure.