Performance Verification
ADA400A Differential Preamplifier
31
Equipment
Required
One digitizing oscilloscope (item 1) to supply power to the ADA400A
One digital multimeter (item 3)
One BNC female to dual banana connector (item 5)
One coaxial cable (item 9)
1.
Set digital multimeter to measure resistance.
2.
Connect banana-to-BNC adapter to digital multimeter.
NOTE
. To avoid an erroneous resistance measurement, make sure the 1 M
Ω
input
resistance jumpers are installed.
3.
Set ADA400A ‘+’ input to
DC
.
4.
Connect coaxial cable from ‘+’ input of ADA400A to the BNC-to-banana
adapter on the multimeter.
5.
Check input resistance = 1 M
Ω
±
2% (980 k
Ω
to
1.02
M
Ω
).
6.
Remove coaxial cable from ADA400A ‘+’ input and connect it to the
‘--’ input.
7.
Set the ADA400A ‘--’ input to
DC
.
8.
Check input resistance = 1 M
Ω
±
2% (980 k
Ω
to
1.02
M
Ω
).
Equipment
Required
One digitizing oscilloscope (item 1)
One dual input coupler (item 6)
One calibration generator (item 4)
One coaxial cable (item 9)
1.
Connect the ADA400A to Ch1 of the oscilloscope and make the equipment
settings according to Table 10.
Table 10: Equipment Settings for Gain Checks
ADA400A Control
Setting
+ Input
DC
-- Input
Gnd
Gain
100
Upper Bandwidth
100 kHz
Input Resistance
Gain
Содержание TEKFAX ADA400A
Страница 4: ......
Страница 8: ...Table of Contents iv ADA400A Differential Preamplifier...
Страница 18: ...Getting Started 6 ADA400A Differential Preamplifier...
Страница 50: ...Performance Verification 38 ADA400A Differential Preamplifier...
Страница 52: ......
Страница 62: ...Adjustment Procedures 50 ADA400A Differential Preamplifier...
Страница 76: ...Replaceable Parts 64 ADA400A Differential Preamplifier...