Adjustment Procedures
48
ADA400A Differential Preamplifier
Table 21: Equipment Settings for CMRR Adjustments (Cont.)
Oscilloscope
Setting
Number of waveforms to average
20
Trigger source
Ch4
Sine Wave Generator
Setting
Output
100 kHz, 1 Vp-p into 50
Ω
2.
Connect a coaxial cable from trigger output of sine wave generator to Ch4 of
oscilloscope.
3.
Connect dual input connector to inputs of ADA400A.
4.
Connect 50
Ω
terminator to dual input connector.
5.
Connect two 10X attenuators to 50
Ω
terminator.
6.
Connect a coaxial cable from sine wave generator output to 10X attenuators.
7.
Adjust trigger level for a steady display.
8.
Adjust amplitude for 5 divisions (10 mV) on screen.
9.
Remove the two 10X attenuators and connect the cable directly to the 50
Ω
terminator.
10.
Set ADA400A ‘--’ input to DC.
11.
Set oscilloscope to 2 mV/div.
12.
Adjust CV4 and R18 for minimum signal amplitude (
≤
1 mV).
13.
Disconnect setup.
Equipment
Required
One digitizing oscilloscope (item 1)
One calibration generator (item 4)
One dual input coupler (item 6)
One 50
Ω
terminator (item 7)
Two coaxial cables (item 9)
One adjustment tool (item 10)
1.
Connect the ADA400A to Channel 1 and make the equipment settings
according to Table 22.
LF Compensation
Adjustments
Содержание TEKFAX ADA400A
Страница 4: ......
Страница 8: ...Table of Contents iv ADA400A Differential Preamplifier...
Страница 18: ...Getting Started 6 ADA400A Differential Preamplifier...
Страница 50: ...Performance Verification 38 ADA400A Differential Preamplifier...
Страница 52: ......
Страница 62: ...Adjustment Procedures 50 ADA400A Differential Preamplifier...
Страница 76: ...Replaceable Parts 64 ADA400A Differential Preamplifier...