Performance Verification
ADA400A Differential Preamplifier
33
12.
Adjust vertical position knob on oscilloscope to center square wave on
screen.
13.
Check DC gain accuracy is
±
2% (4.9 V to 5.1 V).
14.
Set ADA400A gain setting to 0.1.
15.
Set calibration generator to 20 V amplitude.
16.
Adjust vertical position knob on oscilloscope to center square wave on
screen.
17.
Check DC gain accuracy is
±
2% (19.6 V to 20.4 V).
18.
Disconnect setup from the inputs of the ADA400A.
Equipment
Required
One digitizing oscilloscope (item 1)
One dual input coupler (item 6)
1.
Connect the ADA400A to Ch1 of the oscilloscope.
Table 11: Equipment Settings for Offset Adjust
ADA400A
Setting
+ Input
Gnd
-- Input
Gnd
Gain
0.1
Offset
OFF
Coarse
fully CCW
Fine
fully CCW
Oscilloscope
Setting
Vertical input impedance
1
M
Ω
Vertical Coupling
DC
Vertical gain
50 V/div
Vertical Position
0 V (mid-scale)
2.
Verify that the trace is centered vertically on the screen.
3.
Turn the Offset to ON.
4.
Adjust the Coarse knob fully clockwise (CW).
5.
Verify that the oscilloscope trace raises at least 4 divisions.
Offset Adjust
Содержание TEKFAX ADA400A
Страница 4: ......
Страница 8: ...Table of Contents iv ADA400A Differential Preamplifier...
Страница 18: ...Getting Started 6 ADA400A Differential Preamplifier...
Страница 50: ...Performance Verification 38 ADA400A Differential Preamplifier...
Страница 52: ......
Страница 62: ...Adjustment Procedures 50 ADA400A Differential Preamplifier...
Страница 76: ...Replaceable Parts 64 ADA400A Differential Preamplifier...