Adjustment Procedures
44
ADA400A Differential Preamplifier
Equipment
Required
One digitizing oscilloscope (item 1)
One dual input coupler (item 6)
One calibration generator (item 4)
One coaxial cable (item 9)
One adjustment tool (item 10)
1.
Connect the ADA400A to Channel 1 and make the equipment settings
according to Table 18.
Table 18: Equipment Settings for Gain Adjustments
ADA400A
Setting
+ Input
DC
-- Input
Gnd
Gain
100
Upper Bandwidth
100 kHz
Offset
Off
Oscilloscope
Setting
Channel Selection
Ch1
Vertical input impedance
1
M
Ω
Vertical gain
10 mV/div
Time base
200
s/div
Coupling
DC
Vertical Position
0 V (mid-scale)
Trigger
Edge
Trigger position
50%
Trigger source
Ch4
Acquisition mode
Hi Res
Auto Measurement
Amplitude
Bandwidth
Full
Calibration Generator
Setting
Output
50 mV calibrated, 1 kHz
2.
Connect a dual input connector to inputs of ADA400A.
3.
Connect a coaxial cable from the output of the calibration generator to the
dual input connector.
Gain Adjustments