![Tektronix PFG 5105 Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/tektronix/pfg-5105/pfg-5105_instruction-manual_1077779085.webp)
Maintenance
Fuse Replacement -PFG 51 05
Remove t he PFG 5105 from the power module. The
fuses are located under the cover on the right side
of the PFG 5 1 05 (as you face the front of the unit.
The side cover snaps onto the metal rails. Along each
long edge of the cover, there are cutouts about
one-half inch in length. Insert tweezers or a small
straight-edge screwdriver into the cutout near the rear
edge of the cover, and carefully pry the cover away
from the metal rails. Remove the cover.
Three fuses are located toward the rear of the exposed
circuit board, mounted in fuse holders. To remove a
fuse, carefully pull it out of the fuse holder.
4-2
WARNING •
To avoid fire hazard, use only the fuse of correct
type, voltage rating, and current rating
as
specified in the Specification section of this
manual.
After fuse replacement, re-install the side cover, as
follows. Insert the front edge of the cover into the
groove along the front edge of the unit. Then press
the cover down over the rails.