Characteristics
AMPLITUDE (Cont)
Accuracy:
Repeatability:
Transition Time:
Aberrations:
Flatness (See
Fig. 1 -6) :
AMPLITUDE
Range:
Resolution:
Accuracy:
Repeatability:
Table 1-5 (Cont)
OUTPUT CHARACTERISTICS
Performance
Requirements
Supplemental
Information
PULSE (Cont)
±2.5% ±20 mV of programmed value
for 1 .0 to 9.99 V p-p output; ±3% ±5
mV for 10 mV to 999 mV p-p output,
specified for a squarewave output at
1 kHz.
± 1 % for 24 hours.
< 1 5 ns 1 0% to 90% .at full output amp
litude; elsewhere,
<
20 ns, 1 0% to 90%.
< 8% of p-p amplitude ±20 mV from
3.4 V output amplitude to 9.99 V p-p
amplitude.
<
1 0% of p-p amplitude below 3.34 V
p-p output amplitude.
0.5 dB from 0.01 2 Hz to 1 20 kHz,
±2 dB from 1 21 kHz to 1 .2 MHz,
±3 dB from 1 .21 MHz to 12 MHz,
referenced to 1 kHz pulse waveform.
Amplitude
Amplitude set for 5 V pk to pk
(P-P) 7 ,--------------------,
. . . . . . . . . . . . . . . . , · · · · · · · · · . . . . . . · · · · ····· ---,r--
r
.
.
6
4
------ ___________ J
.
�
. . ' .
I
I
I
_ _ _ __ _ _ _ _ _ _ _ _ _ _ _ !
----�------------1
Specified
Flatness
Limits
------------ -----: ______ c
---
c
-
c�-
----
c .
-
�
3 ��-�-�-�-��-�������
1 �
?
1 0
0.1 0.2
0.5
5
1 2
Hz
k z
Frequency (MHz)
Fig. 1 -6. Typical Amplitude Flatness for Pulse.
DC
Specification
±4.99 Vdc from 50Q into 50Q load.
±9.98 Vdc from 50Q into open circuit.
Open circuit values are 2 times the dis
played values.
1 0 mV
±0.6% ±20 mV into 50Q.
± 1 % for 24 hours.
PFG 51 05/5505 Instruction Manual