FRQSTART (Frequency Start)
Command Type:
Setting or query
Setting Syntax:
FRQSTART < freq > [: < units > ]
Examples:
FRQSTART 59.99
FRQSTART 9.99:MHZ
Query Syntax:
FRQSTART?
Query Response:
FRQSTART < freq >
;
FRQSTART 1.0E + 6
;
Discussion:
T his command sets the frequency at which a sweep
will begin. Start frequency must be greater than stop
frequency. Resolution is 4 digits; default unit of meas
ure is Hz.
T he power-on setting is FRQSTART 1 Hz.
Units of measure can be Hz, kHz, or MHz.
FRQSTART and F RQSTOP are not checked for illegal
setting and are not updated until SWEEP ON is sent.
PF G 5105/5505 Instruction Manual
- ------- -------
--
.-,.,
-
Programming
FRQSTOP (Frequency Stop)
Command Type:
Setting or query
Setting Syntax:
F RQST OP < freq > [: < units > ]
Examples:
FRQSTOP 4700
FRQSTOP 11 :MHZ
FRQSTOP 1E6
Query Syntax:
FRQSTOP?
Query Response:
FRQSTOP <freq >
;
FRQSTOP 1.0E + 6
;
Discussion:
T his command sets the frequency at which a sweep
will end. Start frequency must be greater than stop
frequency.
T he resolution is 4 digits
;
the default unit of measure
is Hz. Units of measure can be Hz, kHz, or MHz.
T he power-on setting is FRQSTOP 1200 Hz.
3-1 3