Programming
RNGLCK (Range Lock)
Command Type:
Setting or query
S
ett
ing Syntax:
RNGLCK ON
RNGLCK OFF
Examples:
RNGLCK ON
RNGLCK OFF
Query Syntax:
RNGLCK?
Query Response (one of
t
he following):
RNGLCK ON;
RNGLCK OFF;
Discussion:
The
on
argument turns on the range lock function.
This function limits the generator's frequency output
range to the limits of the frequency range currently
in use (determined by FREQ setting).
When RNGLCK is set
on,
the frequency change is
locked and the output frequency is set to its lowest
value. Changing the frequency parameter will not
affect the range or output frequency. When RNGLCK
is set
on,
the maximum allowable sweep range is
allowed using an external sweep.
The power-on setting is RNGLCK OFF.
For more information, see the discussion of Range
Lock in the Operating Instructions section.
3-20
ROS (Request for Service)
Command Type:
Setting or query
S
ett
ing Syntax:
RQS ON
RQS OFF
Examples:
RQS ON
RQS OFF
Query Syntax:
RQS?
Query Response (one of the following):
RQS ON;
RQS OFF;
Discussion:
The RQS command controls the PFG 5105/5505
service request operation. Valid arguments are:
ON
Allows the PFG 51 05/5505 to generate an
SRQ to report an event or error.
OFF
Prevents the PFG 5105/5505 from generat-
ing an SRQ to report an event or error.
An RQS query returns the current request-for-service
status. The ERR query can be used while RQS is off
to determine if any SRQ conditions have occurred.
See Status and Error Reporting in this section for
more information.
The power-on setting is RQS ON.
PFG 5105/5505 Instruction Manual