Command Type:
Quer
y
Query Syntax:
SEND? < bufnum > [, < bufnum > ] ...
SEND? ALL
Examples:
SEND? 3,5
SEND? 80
SEND? ALL
Query Response:
STOR < butnum >: < binblk > [, < bufnum >: < binblk > ] ... ;
or
STOR ALL: < binblk > ... < binblk > ; (99 settings)
Discussion:
In response to this command, the PF G 5105/5505
transmits over the GPIB, the contents of the stored
settings buffer(s) identified in the argument. Argument
range is 1 to 99. The settings data is sent in binar
y
block format.
PF G 5105/5505 Instruction Manual
Programming
SET?
Command Type:
Quer
y
Query Syntax:
SET?
Query Response Example:
FREQ 1.0E + 3; AMPL 5.0; OFFS 0; DC 0; RATE
1 0.0E-6:S; NBUR 2; F RQSTART 1.0; FRQSTOP
1.2E + 3; SWEEP OFF ; FUNC SINE; MODE CONT;
TRIG MANUAL; AM OFF ; FM OFF ; OUT OFF ; FRQL
ON; RNGLCK OFF ; OT OFF ; ROS ON; USER OFF ;
DELAY 100E-9; DCYCLE O; PRELEVEL ECL; DISP
F REQUENCY; WIDTH 0.04E-6;
Discussion:
This command returns the status of all instrument
settings that can be set and that respond to a quer
y
command with the following exception:
stored front panel settings
3-21