Programming
Command Type:
Setting or query
Setting Syntax:
MODE CONT
MODE BURST
Examples:
MODE CONT
MODE BURST
Query Syntax:
MODE?
MODE
MODE TRIG
MODE GATE
MODE TRIG
MODE GATE
MODE SYNT
MODE SYNT
Query Response (one of the following):
MODE CONT;
MODE TRIG;
MODE BURST;
MODE GATE; MODE SYNT;
Discussion:
This command sets the trigger mode to the mode
specified by the argument.
CONT
Sets the PFG 5105/5505 to generate a
continuous output signal. Trigger events
are ignored. CONT is the power-on setting.
TRIG
Sets the PFG 5105/5505 to the triggered
output mode. One cycle of the output
signal is generated for each trigger event.
Trigger sources include an ex
t
ernal trig
ger, internal trigger, manual trigger, or
Group Execute Trigger < GET > over the
GPIB. See also the TRIG and DT command
descriptions.
BURST
Sets the PFG 5105/5505 to the burst mode.
When a trigger occurs, the instrument pro
duces a burst of the programmed output
signal; the number of cycles is determined
by the N BURST parameter. All trigger
sources apply.
GATE
Sets the PFG 5105/5505 to the gated mode.
Output is generated while the MAN key is
pressed or the trigger/gate input is enabled.
If the gate signal on the trigger input is
removed in the middle of a cycle, the cycle
is completed. < GET > toggles the gate
on or off when DT is set to GATE.
SYNT
This mode is available only on an Option
02 instrument. It sets the PFG 5105/5505
to the synthesizer mode. If this mode is
selected and the instrument does not have
the option installed, an execution error is
reported.
NBURST (Number of Burst Cycles)
Command Type:
Set
t
ing or query
Setting Syntax:
NBURST < # cycles >
Examples:
NBURST 9999
NBURST 10
Query Syntax:
NBURST?
Query Response:
NBURST < # cycles >;
Discussion:
This command sets the number of cycles that will
be output in burst mode. The argument must be a
whole number. The power-on set
t
ing is 2 cycles; the
range is 1 to 9999.
3-1 6