Power-On/Initial Conditions
When powered on, the PFG 51 05/5505 mi
c
ropro
c
es
sor performs a diagnosti
c
routine (self test) to
c
he
c
k
the fun
c
tionality of the ROM and RAM. If no ROM or
RAM error is found, the mi
c
ropro
c
essor goes on to
c
he
c
k the fun
c
tionality of the other instrument hard
ware. When the self test is finished, the instrument
enters the lo
c
al state (LOCS) with the following default
settings:
Table 2-1
POWER-ON DEFAULT SETTINGS
AM
OFF
AMPLITUDE
5 V
DC
0 V
DCYCLE
0
DELAY
0
DT
OFF
FM
OFF
FRQLCK
ON
FRQSTART
1 Hz
FRQSTOP
1 200 Hz
FREQUENCY
1 kHz
FUNCTION
SINE
MODE
CONT
NBURST
2
OFFSET
O V
OUTPUT
OFF
PERIOD
1 ms
RATE
10
µs
RNGLCK
OFF
RQS
ON
SWEEP
OFF
TRIG
MAN
USEREQ
OFF
WIDTH
0.5 ms
The GPIB address and termination are fa
c
tory-set to
8 and LF with EOI, respe
c
tively. New values for these
parameters
c
an be entered and stored using SPCL
240 and 241. See Spe
c
ial Fun
c
tions. later in this
se
c
tion, for more information.
The SRQ line on the GPIB is also asse
rt
ed unless
the GPIB address is set to 31 (ignore GPIB
c
om
mands). If the instrument is polled by the
c
ontroller,
PFG 51 05/5505 Instru
c
tion Manual
Operating Instructions
the status byte returned will be "0100 0001 " (65
de
c
imal; power-on SRQ).
If a ROM or RAM error is found, an error
c
ode is
displayed in the front-panel display. The instrument
will not respond to input from the front panel or the
GPIB interfa
c
e. Internal errors dete
c
ted after the ROM
and RAM tests have passed will be repo
rt
ed to both
the front panel and the GPIB. The instrument will
respond to input and attempt to fun
c
tion in spite of
the internal error. An error
c
ode may be removed
from the display by pressing the INST ID key, by
sta
rt
ing a numeri
c
entry, by in
c
rementing the sele
c
ted
parameter, by pressing the
c
lear key, or by a transition
into the remote state (REMS).
GPIB Indicators
ADRS
REMOTE
Illuminates when the PFG 51 05/5505 is
addressed as a listener or as a talker
over the GPIB.
Illuminates when the PFG 51 05/5505 is
operating under remote program
c
ontrol
via the GPIB.
CONTROLS AND CONNECTORS
The following des
c
ription explains the fun
c
tion of ea
c
h
front panel
c
ontrol and
c
onne
c
tor. The number asso
c
iated with ea
c
h des
c
ription pinpoints the lo
c
ation of
the key or
c
onne
c
tor on the front panel in Fig. 2-5.
G)
PARAMETER KEYS
The PARAMETER keys are used to sele
c
t a parameter
for the waveform sele
c
ted for output from the
PFG 51 05/5505; to display the
c
urrent value of th�
parameter; and to
c
hange its value and/or units of
measure using the DATA keypad. The PARAMETER
keys are mutually ex
c
lusive in operation.
FREQ
AMPL
Sele
c
ts the FREQuen
c
y parameter and
displays the
c
urrent frequen
c
y of the
output fun
c
tion and its
c
urrent amplitude.
Frequen
c
y
c
an be
c
hanged via front
panel keys. Be
c
omes period entry and
display when SPCL 21 O is used.
Sele
c
ts the AMPLitude parameter and
displays the
c
urrent amplitude of the
output fun
c
tion. The amplitude
c
an be
c
hanged via front panel keys. Amplitude
and offset are both displayed when either
parameter is sele
c
ted.
2
-5