Triggering
3- 120
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Overview
Control elements and resources
To trigger on a sequence (Cont.)
To trigger on
B events
1.
To set the time base to trigger after an A trigger and a
specified number of B trigger events, from the toolbar,
touch Trig, and select the A-->B Seq tab of the Trigger
control window.
2.
Touch A Then B Trig on nth Event.
3.
To set the number of B trigger events, touch Trig Event,
and use the multipurpose knob, keyboard, or up and
down arrows to set the number of events.
4.
To set the B trigger level, touch B Trig Level, and use
the multipurpose knob or keypad to set the level.
Содержание CSA7154
Страница 4: ......
Страница 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 224: ...Displaying Waveforms 3 146 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 366: ...Accessing Online Help 3 288 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 408: ...Appendix A Specifications A 42 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 422: ...Appendix B Automatic Measurements Supported B 14 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...