Related control elements and resources
To connect the probe calibration fixture (Cont.)
P6249 or P7240
Connect the probe tip to the short pin and the probe ground to
the long pin as shown.
P6246, P6247, P6248, or P7330
Connect the probe + input to the signal pin
and the probe -- input to the ground pin as
shown. There is no connection to the probe
ground input.
TCP-202
Snap the current probe closed around the current loop as
shown. Be sure to align the polarity arrows on the current
probe and the fixture.
CT-6
After you are finished with this probe, be sure to
replace the small jumper.
Remove the small jumper on the A side of the
fixture. Slip the CT-6 probe head over the outboard
pin, and then replace the jumper.
Содержание CSA7154
Страница 4: ......
Страница 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 224: ...Displaying Waveforms 3 146 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 366: ...Accessing Online Help 3 288 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 408: ...Appendix A Specifications A 42 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 422: ...Appendix B Automatic Measurements Supported B 14 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...