Measuring Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 179
Overview
Related control elements and resources
To calibrate probes (Cont.)
Check the
calibration
status
11.
From the toolbar, touch the
VERT
button to display the
instrument Vertical setup control window.
12.
Touch the Probe
Cal
button to display the vertical Probe
Cal control window.
13.
Select the instrument channel to which the probe is
attached.
14.
Check the Probe Status readout. See right.
H
Initialized. Indicates that the probe has not been
calibrated on the selected channel.
H
Pass. Indicates that the probe has been calibrated
on the selected channel.
Note.
Probes without a TekProbe interface cannot be
calibrated.
For further
assistance
15.
Touch the Help button to access the online assistance.
See page 3--283 to learn about using online help.
Содержание CSA7154
Страница 4: ......
Страница 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 224: ...Displaying Waveforms 3 146 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 366: ...Accessing Online Help 3 288 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 408: ...Appendix A Specifications A 42 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 422: ...Appendix B Automatic Measurements Supported B 14 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...