Measuring Waveforms
3- 182
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Overview
Related control elements and resources
To deskew channels (Cont.)
Compensate
probe timing
(deskew)
2.
Connect fixture to the instrument (see
To Connect the
Probe Calibration Fixture
on page 3--173). CSA7000
Series, TDS7000 Series, and TDS6404: The jumper
must be installed.
3.
Connect up to four probes to the fixture.
4.
Display all the channels that you want to deskew.
5.
Push the
AUTOSET
button on the instrument.
6.
Adjust the Vertical
SCALE
and
POSITION
for each
channel so that the signals overlap and are centered on
screen.
7.
Adjust the horizontal
POSITION
, as indicated by the red
pointer at the top of the graticule area, to the center of
the display. The red pointer is the trigger position and all
channels will be aligned to this point.
8.
Adjust the horizontal
SCALE
so that the differences in
the channel delays are clearly visible.
9.
From the toolbar, touch the
VERT
button to display the
Vertical setup control window.
10.
Touch the Probe
Deskew
button to display the
channel-deskew control window.
11.
Select one of the channels. The channel that is the
trigger source will already be at the trigger position and
should not need to be deskewed.
Note.
Do the next step at a signal amplitude within the
same attenuator range (vertical scale) as your planned
signal measurements. Any change to the vertical scale
after deskew is complete may introduce a new
attenuation level (you can generally hear attenuator
settings change) and, therefore, a slightly different
signal path. This different path may cause a variation in
timing accuracy between channels.
Содержание CSA7154
Страница 4: ......
Страница 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 224: ...Displaying Waveforms 3 146 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 366: ...Accessing Online Help 3 288 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 408: ...Appendix A Specifications A 42 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 422: ...Appendix B Automatic Measurements Supported B 14 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Страница 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...