Operating
Instructions—
Type
3B5
triggering
point
may
be
enough
so
it
is
impossible
to
main
tain
a
stable
display. In
such
cases,
use
DC coupling.
DC.
DC
trigger
coupling
is
available
only
when
the
sweep
is
triggered from
an
external
signal
in
the
EXT
posi
tion.
DC
coupling
can
be
used
to
provide
stable
triggering
with
low-frequency signals which
would
be
attenuated
in
the
AC
position
or with
low-repetition
rate
signals.
The
LEVEL
control
can
be
adjusted
to
provide
triggering
at
the
desired
DC
level
on
the
waveform.
Trigger
Slope
The
Trigger
SLOPE
lights
indicate
the portion
of
the
trigger
signal
to
which
the
trigger
circuits respond.
When
the
minus
(
—
)
light
is
on, the
display
starts
on
the
negative-going
portion
of
the
waveform;
the
plus
(+)
light
indicates
that
the
display starts
on
the positive-going
portion
of
the
wave
form.
The
trigger
slope
can be
changed
by pressing
the
CHANGE
SLOPE
button
(or by
external
program).
Each
time
the
button
is
pressed
the
trigger
slope
changes
from
(
—
)
to
(+)
or
vice
versa.
The
operation
of
the
CHANGE
SLOPE
switch
in
conjunction
with
the
LEVEL
control
is
shown
in
Fig.
2-3.
When
several
cycles
of
a
signal
are
displayed,
the trigger
slope
is
often
unimportant.
However,
if
only
a
certain
portion
of
a
cycle is
to
be
displayed,
correct
slope
setting
is
important
to
provide
a
display
which
starts
at
the desired
point
on
the
input
signal.
Trigger
Level
The
front-panel
Trigger LEVEL control
(or
external
program
level
control)
selects
the
voltage
level
on
the
triggering
waveform
at
which
the
sweep is
triggered.
When
the
LEVEL
control
is
set
in
the
clockwise
region,
the trigger circuit
responds
at
a
more
positive
point
on
the trigger
signal.
When
the
LEVEL
control is in
the
counterclockwise
region,
the
trigger
circuits
responds at
a
more
negative
point
on
the
trigger
signal.
Fig.
2-3
illustrates
this
effect
for
both
trigger
slope
conditions.
The
NOT
TRIG'D
readout
on
the readout panel
provides
a
convenient
indication
of
the
condition
of
the
trigger
circuit.
If
the
LEVEL
control
is correctly
adjusted
with
an
adequate
trigger
signal
applied,
the
readout is
off.
However,
if
the
LEVEL
control
is
misadjusted, the
Trigger
Function
switch
incorrectly
set,
or
the
trigger
signal
too
low
in
amplitude,
the
NOT TRIG'D
readout
lights
up.
This
readout
feature
can
be
used
for
a
general
indication
of
correct
triggering
and
is
particularly
useful
when
setting
up
the
trigger
circuits
when
a
trigger
signal
is
available
without
a
trace
displayed
on
the
CRT
(such as
single
sweep
operation).
To
set
the
LEVEL
control,
first
select
the
trigger
source,
coupling
and
slope. Then
set
the
LEVEL
control
to
midrange
(except
for DC
coupling).
If
the display
does
not
start
at
the
desired
point,
adjust
the
LEVEL
control
for
correct
trig
gering.
With
DC
coupling
(external
trigger
only),
stable
triggering
may
be
obtained
at
any
setting
of
the
LEVEL
control as
determined
by
the
DC
level
of
the
trigger
signal.
To
obtain
correct
DC
trigger
level,
turn
the
LEVEL
control
counterclockwise.
Then,
turn
the
LEVEL
control
clockwise
until
the
display is
triggered
at
the
correct
DC
level
on
the
displayed
signal.
Auto
Triggering
The
sweep
is automatically
triggered
in
the
INT-AUTO
position
of
the
Trigger Function
switch (or
when
selected
by
external
program).
Auto
triggering
can
be
used
for most
applications.
This
trigger
mode
provides
a
stable
display
when
a
trigger
signal
is
available
(above
about
50 Hz).
Complete
rotation
of
the
LEVEL
control
provides
a
range
of
adjustment
between
about
10%
and
90%
of
the
peak-to-
peak
level
of
the
triggering slope.
The
front-panel
LEVEL
control also
determines
the
auto
triggering
level for
External
Mode
operation.
The
trigger circuit automatically
compen
sates
for
differences
in
trigger
signal
amplitude
to
provide
this
range
of
LEVEL
control
adjustment
for
all
signals
which
meet
the
minimum
amplitude
requirements.
When
the
LEVEL
control
is
set for
the
desired
triggering
point
at
a
certain
percentage
of
the
leading
edge, it triggers
at
this
same
point
even
if
the
signal
amplitude
is
increased
(e.g.,
if
set
to
trigger
at the
30%
point
on
a
one-division
signal,
it
also
triggers
at
the
30%
point
if
the
amplitude
is
increased
to
eight
divisions;
see
Fig. 2-4).
NOTE
The
auto
trigger
mode may
not
produce
the
desired
display
of
complex signals
such
as
pulse
trains
near
the
low-frequency
limit
of
this
mode
(about
50
hertz).
The
triggering
point
may
shift
from
the
desired
pulse to
a
lower
amplitude
pulse
which
follows.
The
resultant
display
will
appear
unstable.
In
such
cases,
use
normal internal
trig
gering
or
external
triggering.
The
auto
trigger
mode
is particularly useful
when
a
reference trace is
needed
and
no
trigger
signal is
available.
Then,
the
Sweep
Generator
free
runs
to produce
a
reference
trace.
The
trace also
free
runs
when
the
repetition
rate
of
the
trigger
signal
drops
below
about
50
Hz.
When
a
trigger
signal
is again
present,
the
free
running
condition
ends
and
the
sweep
returns
to
a
triggered,
stable
display.
When
a
seek
command
is
received
in
the
INT-AC,
EXT-
AC,
EXT-DC
positions
of
the
Trigger
Function
switch,
the
trigger
circuits automatically
switch
to
auto
triggering
if
a
trigger
signal
is
not
present
or
if
the
LEVEL control
is
mis
adjusted.
Then,
a
stable
auto
triggered
display
is
presented,
if
possible.
This
feature
assures
that
a
stable
display
is
presented,
whenever
possible,
in
the
Seek
Mode
of
operation.
Selecting
Sweep
Rate
The
MANUAL
TIME/DIV
switch
selects
the
calibrated
sweep
rate
for
the
Sweep Generator in
the
Manual
Mode
of
operation.
The
VARIABLE
control
provides
continuously
variable
sweep
rates
between
the
settings
of
the
MANUAL
TIME/DIV
switch.
Whenever
the
UNCAL
readout
is
on,
the
sweep
rate
is
uncalibrated.
The
readout
is off
when
the
VARIABLE control
is
set
to
the
CAL
position.
When
making
time measurements from
the
graticule,
the
area
between
the
first
and
ninth
graticule
lines
provides
the
most linear
time
measurements
(see
Fig.
2-5).
Therefore,
the
first
and
last
division
of the
display
should
not
be
used
for making
accurate
time
measurements.
Position
the
start
of
the
timing
area
to
the
first
graticule
line
and
set
the
MANUAL
TIME/DIV
switch
so
the
end
of
the
timing
area
falls between
the
first
and
ninth
graticule
lines.
2-8
Содержание 3B5
Страница 4: ...Fig 1 1 Type 3B5 Automatic Programmable Time Base unit Type 3B5...
Страница 15: ...Operating Instructions Type 3B5 TYPE 3B5 CONTROL SET UP CHART Fig 2 2 Control set up chart 2 7...
Страница 48: ...CO I o Fig 3 13 Delay and Timing Circuit logic block diagram Circuit Description Type 3B5...
Страница 61: ...GO i GO GO Fig 3 22 Seek Ciicuit Logic block diagram Circuit Description Type 3B5...
Страница 70: ...u k KJ Fig 3 29 Circuit conditions for Manual Mode operation Circuit Description Type 3B5...
Страница 71: ...w K w Fig 3 30 Circuit conditions for Seek Mode operation Circuit Description Type 3B5...
Страница 72: ...w I u U Fig 3 31 Circuit condition for External Mode operation Circuit Description Type 3B5...
Страница 88: ...Maintenance Type 3B5 Fig 4 9 Location of components on Logic Card 4 14...
Страница 89: ...u Oi Fig 4 10 Location of components on Counter Card Maintenance Type 3B5...
Страница 92: ...NOTES I...
Страница 104: ...NOTES...
Страница 106: ...Calibration Type 3B5 Fig 6 1 Recommended calibration equipment...
Страница 160: ......
Страница 176: ...J400 RtADOUT BOARD 3B5 PLUG IN A READOUT...
Страница 182: ...397 R E A D O U T B O A R D 10 6b READOUT BOARD...
Страница 184: ...FIG 1 FRONT SWITCHES TYPE 3B5 AUTOMATIC PROGRAMMABLE TIME BASE...
Страница 185: ...FIG 2 CHASSIS REAR 3 GS to TYPE 3B5 AUTOMATIC PROGRAMMABLE TIME BASE...
Страница 186: ...OPTIONAL ACCESSORIES...