1705A Spectrum Monitor
5- 1
Section 5
Checks And Adjustments
This section consists of two separate procedures. The first, a Performance
Check, is used to determine compliance with the Performance Requirements in
the Specification. The second is the Adjustment Procedure, that provides the
instructions on how to adjust the instrument and return it to operation within the
specification.
In both procedures, front- and rear-panel controls and connectors, on the
instrument under test, are fully capitalized (e.g., 300 kHz RESOLUTION).
Control and connector names on test equipment and internal controls and
adjustments for the instrument under test are initial capitalized (e.g., Time/Div,
Geometry, etc.).
Limits, tolerances, and waveforms given in this section are guides to adjustments
and checks. They are not instrument specifications, except when listed in the
Performance Requirement column of the Specification Tables in Section 1 of this
manual.
Recommended Equipment List
The following equipment and accessory items are required to do the Performance
Check and/or Adjustment Procedures. Broad specifications are followed by a
piece of equipment that meets these specifications; in most cases, the recom-
mended instrument was used in preparing the procedures that follow.
1.
Test Oscilloscope
Vertical Amplifier:
30 MHz Bandwidth, 1 mV Sensitivity.
Time Base:
10 ns/Division to 5 ms/Division sweep speeds, triggering to 5 MHz.
For example: a TEKTRONIX 7603 Oscilloscope with a 7A18 Dual-Trace
Amplifier and a 7B53A Dual Time Base, or a TEKTRONIX 11403A
Oscilloscope with a 11A34V Video Amplifier and an 11T5H Video Trigger.
Also a 10X probe, P6106 (Tektronix Part No. 010-6106-03).
Electrical Instruments
Содержание 1705A
Страница 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Страница 17: ...Preface xvi 1705A Spectrum Monitor ...
Страница 18: ...Introduction ...
Страница 19: ......
Страница 38: ...Operating Instructions ...
Страница 39: ......
Страница 61: ......
Страница 62: ...Installation ...
Страница 63: ......
Страница 72: ...Theory of Operation ...
Страница 73: ......
Страница 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Страница 94: ...Checks and Adjustments ...
Страница 95: ......
Страница 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Страница 124: ...Maintenance ...
Страница 125: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...
Страница 154: ...Options ...
Страница 155: ......
Страница 158: ...Replaceable Electrical Parts ...
Страница 159: ......
Страница 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Страница 190: ...Diagrams Circuit Board Illustrations ...
Страница 191: ......
Страница 193: ......
Страница 213: ......
Страница 214: ...Replaceable Mechanical Parts ...
Страница 215: ......
Страница 223: ......