Maintenance
6- 2
1705A Spectrum Monitor
detergent and water. Abrasive cleaners should not be used. If the circuit board
assemblies must be removed for cleaning, follow the instructions for removal/re-
placement under the heading of Corrective Maintenance.
After cleaning, allow the interior to thoroughly dry before applying power to the
instrument.
CAUTION.
Do not allow water to get inside any enclosed assembly or component. Do
not clean any plastic materials with organic cleaning solvents, such as benzene,
toluene, xylene, acetone, or similar compounds, because they may damage the plastic.
After cleaning, carefully check the instrument for defective connections,
damaged parts, and improperly seated transistors or integrated circuits. The
remedy for most visible defects is obvious; however, if heat--damaged parts are
discovered, determine the cause of overheating before replacing the damaged
part, to prevent additional damage.
Periodic checks of the transistors and integrated circuits are not recommended.
The best measure of performance is the actual operation of the component in the
circuit.
This instrument contains electrical components that are susceptible to damage
from static discharge. Static voltages 1 kV to 30 kV are common in unprotected
environments. Table 6--1 shows the relative static discharge susceptibility of
various semiconductor classes.
Table 6- 1: Static Susceptibility
Relative Susceptibility Levels
Voltage
1
CMOS
100 V -- 500 V
2
ECL
200 V -- 500 V
3
SCHOTTKY SIGNAL
DIODES
250 V
4
SCHOTTKY TTL
500 V
5
HF BIPOLAR TRAN-
SISTORS
400 to 600 V
6
JFETS
600 to 800 V
7
LINEAR
CIRCUITS
400 to 1000 V est.
8
LOW POWER SCHOTT-
KY TTL
900 V
9
TTL
1200 V
Visual Inspection
Static- Sensitive
Components
Содержание 1705A
Страница 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Страница 17: ...Preface xvi 1705A Spectrum Monitor ...
Страница 18: ...Introduction ...
Страница 19: ......
Страница 38: ...Operating Instructions ...
Страница 39: ......
Страница 61: ......
Страница 62: ...Installation ...
Страница 63: ......
Страница 72: ...Theory of Operation ...
Страница 73: ......
Страница 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Страница 94: ...Checks and Adjustments ...
Страница 95: ......
Страница 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Страница 124: ...Maintenance ...
Страница 125: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...
Страница 154: ...Options ...
Страница 155: ......
Страница 158: ...Replaceable Electrical Parts ...
Страница 159: ......
Страница 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Страница 190: ...Diagrams Circuit Board Illustrations ...
Страница 191: ......
Страница 193: ......
Страница 213: ......
Страница 214: ...Replaceable Mechanical Parts ...
Страница 215: ......
Страница 223: ......