Checks and Adjustments
1705A Spectrum Monitor
5- 9
!
!
SG503
VARIABLE
AUTOTRANSFORMER
70 MHz
INPUT
1705A
Figure 5-3: Output of the SG503 connected directly to the 1705A 70 MHz INPUT
g.
Set the leveled sine wave generator Frequency to 45 MHz, Amplitude
Multiplier to .01, and Amplitude to 0.5 (--42 dBm).
h.
Set the 1705A SPAN/DIV to FULL, and the CENTER FREQUENCY to
read 45 MHz. Check to see that the marker is intensified.
i.
Set the 1705A SPAN/DIV to 1 MHz and check that the marker is on
screen.
j.
Position the marker under the readout caret with the CENTER FRE-
QUENCY control.
k. CHECK
-- that readout reads 45,
±
1 count.
l.
Set SPAN/DIV to 100 kHz.
m. CHECK
-- that the marker is on screen.
n.
Repeat steps h. through m., substituting 95 MHz.
o.
Leave SPAN/DIV at 100 kHz.
Содержание 1705A
Страница 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Страница 17: ...Preface xvi 1705A Spectrum Monitor ...
Страница 18: ...Introduction ...
Страница 19: ......
Страница 38: ...Operating Instructions ...
Страница 39: ......
Страница 61: ......
Страница 62: ...Installation ...
Страница 63: ......
Страница 72: ...Theory of Operation ...
Страница 73: ......
Страница 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Страница 94: ...Checks and Adjustments ...
Страница 95: ......
Страница 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Страница 124: ...Maintenance ...
Страница 125: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...
Страница 154: ...Options ...
Страница 155: ......
Страница 158: ...Replaceable Electrical Parts ...
Страница 159: ......
Страница 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Страница 190: ...Diagrams Circuit Board Illustrations ...
Страница 191: ......
Страница 193: ......
Страница 213: ......
Страница 214: ...Replaceable Mechanical Parts ...
Страница 215: ......
Страница 223: ......