Theory of Operation
1705A Spectrum Monitor
4- 19
Q5 is a current amplifier feeding the output stage. Q3 and Q4 form a push-pull
output stage. Q3 acts as a 2.7 mA constant current pull-up, while Q4 is the
pull-down transistor. C6 speeds up the amplifier by coupling ac signals to the
base of Q3. CR2 and R41 protect the amplifier during crt arcing.
The pinout for the CRT is shown in Figure 4-3.
Pin
Description
1
Filament (f)
2
Cathode (k)
3
GRID (g1)
4
FOCUS (g3)
5
ASTIG (g4)
6
GEOM (g5)
7
VERT PLATE (y2)
9
VERT PLATE (y1)
11
HORIZ PLATE (x2)
12
1st ANODE (g2)
13
HORIZ PLATE (x1)
14
Filament (f)
14
13
12
11
9
7
6
5
4
3
2
1
Figure 4-3: Pinout of the CRT Socket
CRT
Содержание 1705A
Страница 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Страница 17: ...Preface xvi 1705A Spectrum Monitor ...
Страница 18: ...Introduction ...
Страница 19: ......
Страница 38: ...Operating Instructions ...
Страница 39: ......
Страница 61: ......
Страница 62: ...Installation ...
Страница 63: ......
Страница 72: ...Theory of Operation ...
Страница 73: ......
Страница 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Страница 94: ...Checks and Adjustments ...
Страница 95: ......
Страница 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Страница 124: ...Maintenance ...
Страница 125: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...
Страница 154: ...Options ...
Страница 155: ......
Страница 158: ...Replaceable Electrical Parts ...
Страница 159: ......
Страница 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Страница 190: ...Diagrams Circuit Board Illustrations ...
Страница 191: ......
Страница 193: ......
Страница 213: ......
Страница 214: ...Replaceable Mechanical Parts ...
Страница 215: ......
Страница 223: ......