Maintenance
1705A Spectrum Monitor
6- 7
PRESS [INPUT] KEY TO EXIT
ERROR : CANNOT READ OR WRITE
TO 2444
Figure 6-4: 1705A power up error message
When this message is displayed, any subsequent measurements (if possible)
must be considered suspect.
The DAC test is used to test the readout capability, crt geometry and focus, and
provides a convenient method of determining if Trace Rotation needs to be
adjusted.
Figure 6-5 shows the waveform superimposed over the crt graticule to demon-
strate its value as an adjustment tool. The DAC test is selected, from the Test
Auxiliary menu, by pushing the 1705A RESOLUTION push button. The test in
progress is exited any time the INPUT push button is pushed.
DAC Test
Содержание 1705A
Страница 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Страница 17: ...Preface xvi 1705A Spectrum Monitor ...
Страница 18: ...Introduction ...
Страница 19: ......
Страница 38: ...Operating Instructions ...
Страница 39: ......
Страница 61: ......
Страница 62: ...Installation ...
Страница 63: ......
Страница 72: ...Theory of Operation ...
Страница 73: ......
Страница 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Страница 94: ...Checks and Adjustments ...
Страница 95: ......
Страница 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Страница 124: ...Maintenance ...
Страница 125: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...
Страница 154: ...Options ...
Страница 155: ......
Страница 158: ...Replaceable Electrical Parts ...
Страница 159: ......
Страница 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Страница 190: ...Diagrams Circuit Board Illustrations ...
Страница 191: ......
Страница 193: ......
Страница 213: ......
Страница 214: ...Replaceable Mechanical Parts ...
Страница 215: ......
Страница 223: ......