Checks and Adjustments
5- 4
1705A Spectrum Monitor
6.
Check Resolution Filter
b.
CHECK
-- that the marker width 0.6 Divisions down from the top is
3 Divisions,
±
1 Division.
e.
CHECK
-- that the marker width 0.6 Divisions down from the top of the
signal is
≤
2 minor Divisions.
f.
CHECK
-- that the 10 kHz marker amplitude matches the highest point
of the 300 kHz marker amplitude within 1 minor Division.
7.
Check 70 MHz Gain and Flatness
e.
CHECK
-- for a marker amplitude change of 2 Divisions,
±
2 minor
Divisions when switching the leveled sine wave generator Amplitude
Multiplier between .01 and 0.1 (a 20 dB change).
i.
CHECK
-- that the tip of the marker is on the crt center line,
±
1 minor
Division (vertically).
8.
Check Video Filter
d.
CHECK
-- that the baseline noise amplitude drops approximately 50%
when the VIDEO FILTER is ON.
9.
Check Sweep Speed
b.
CHECK
-- for a solid trace with almost no flicker.
d.
CHECK
-- for approximately 3 to 4 sweeps per second.
Alternate Method
e.
CHECK
-- for a ramp duration of 20 ms
±
10 ms.
h.
CHECK
-- for a ramp duration of 200 ms
±
100 ms.
10.
Check L--Band Linearity
e.
CHECK
-- for 10 frequency markers.
f.
CHECK
-- that each marker is within 1 minor Division of a major
graticule line.
11.
Check L--Band SPAN/DIV and Readout
d.
CHECK
-- for 1 marker every 5 Divisions,
±
1 Division.
i.
CHECK
-- for 1 marker every 5 Divisions,
±
1 Division.
o.
CHECK
-- that the marker is intensified.
q.
CHECK
-- that the marker is on screen.
Содержание 1705A
Страница 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Страница 17: ...Preface xvi 1705A Spectrum Monitor ...
Страница 18: ...Introduction ...
Страница 19: ......
Страница 38: ...Operating Instructions ...
Страница 39: ......
Страница 61: ......
Страница 62: ...Installation ...
Страница 63: ......
Страница 72: ...Theory of Operation ...
Страница 73: ......
Страница 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Страница 94: ...Checks and Adjustments ...
Страница 95: ......
Страница 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Страница 124: ...Maintenance ...
Страница 125: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...
Страница 154: ...Options ...
Страница 155: ......
Страница 158: ...Replaceable Electrical Parts ...
Страница 159: ......
Страница 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Страница 190: ...Diagrams Circuit Board Illustrations ...
Страница 191: ......
Страница 193: ......
Страница 213: ......
Страница 214: ...Replaceable Mechanical Parts ...
Страница 215: ......
Страница 223: ......