Maintenance
1705A Spectrum Monitor
6- 3
Observe the following precautions to avoid damage:
1.
Minimize handling of static-sensitive components.
2.
Transport and store static-sensitive components or assemblies in their
original containers, on a metal rail, or on conductive foam. Label any
package that contains static-sensitive components or assemblies.
3.
Discharge the static voltage from your body, by wearing a wrist grounding
strap, while handling these components. Servicing static-sensitive assem-
blies or components should be done only at a static-free work station by
qualified personnel.
4.
Nothing capable of generating or holding a static charge should be allowed
on the work station surface.
5.
Keep the component leads shorted together whenever possible.
6.
Pick up the components by the body, never by the leads.
7.
Do not slide the components over any surface.
8.
Avoid handling components in areas that have a floor or work surface
covering capable of generating a static charge.
9.
Use a soldering iron that is connected to earth ground.
10.
Use only special antistatic, suction, or wick-type desoldering tools.
Instrument performance should be checked after each 2000 hours of operation, or
every 12 months, if used intermittently. This will help to ensure maximum
performance and assist in locating defects that may not be apparent during
regular operation. The Performance Check Procedure and the Adjustment
Procedure are in Section 5.
TROUBLESHOOTING
The 1705A is equipped with a microprocessor and, as with all of the new
Tektronix monitors, it is capable of some power-up and off-line diagnostics. A
thorough understanding of the information available in the diagnostic routines
will assist in isolating problems and speed repairs.
Performance Checks and
Readjustments
Содержание 1705A
Страница 13: ...Service Safety Summary xii 1705A Spectrum Monitor ...
Страница 17: ...Preface xvi 1705A Spectrum Monitor ...
Страница 18: ...Introduction ...
Страница 19: ......
Страница 38: ...Operating Instructions ...
Страница 39: ......
Страница 61: ......
Страница 62: ...Installation ...
Страница 63: ......
Страница 72: ...Theory of Operation ...
Страница 73: ......
Страница 93: ...Theory of Operation 4 20 1705A Spectrum Monitor ...
Страница 94: ...Checks and Adjustments ...
Страница 95: ......
Страница 123: ...Checks and Adjustments 5 28 1705A Spectrum Monitor ...
Страница 124: ...Maintenance ...
Страница 125: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...
Страница 154: ...Options ...
Страница 155: ......
Страница 158: ...Replaceable Electrical Parts ...
Страница 159: ......
Страница 189: ...Replaceable Electrical Parts 8 30 1705A Spectrum Monitor ...
Страница 190: ...Diagrams Circuit Board Illustrations ...
Страница 191: ......
Страница 193: ......
Страница 213: ......
Страница 214: ...Replaceable Mechanical Parts ...
Страница 215: ......
Страница 223: ......